User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: FLUKE 8600A PERFORMANCE TESTS DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 43 NUMBER OF LINES: 92 CONFIGURATION: Fluke 5100B w/Wideband Fluke 5205A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Tests should be conducted under the following conditions 1.001 DISP AMBIENT TEMPERATURE 23 +/-5 DEGREES CELSIUS, RELATIVE 1.001 DISP HUMIDITY <80% 1.002 DISP TURN THE INSTRUMENT ON 1.003 DISP CONNECT TEST CABLE TO V-OHM AND LO INPUT TERMINALS 1.004 HEAD {DC PERFORMANCE TEST} 1.005 5100 200 190mV 0.1U 2W 2.001 5100 200 -19mV 0.03U 2W 3.001 5100 2 1.9V 0.0005U 2W 4.001 5100 2 -1.9V 0.0005U 2W 5.001 5100 2 0.9V 0.0003U 2W 6.001 5100 20 19V 0.005U 2W 7.001 5100 20 -19V 0.005U 2W 8.001 5100 200 190V 0.05U 2W 9.001 5205 1200 1000V .3U E 2W 10.001 5100 A -0.17V 9e-005U 2W 11.001 5100 A -1.7V 0.0004U 2W 12.001 5100 A -17V 0.004U 2W 13.001 5100 A -170V 0.04U 2W 14.001 5205 A -1000V .3U E 2W 15.001 HEAD {OHMS PERFORMANCE TEST} 15.002 DISP SELECT 8600A OHMS FUNCTION WITH AUTORANGE AND 15.002 DISP SHORT THE V-OHM AND LO INPUT TERMINALS 15.003 EVAL {Does the display read 00.00 to 00.03?} 16.001 DISP REMOVE SHORT 16.002 RESF 200 100Z 0.13U 2W 17.001 RESF 2 1KZ 0.0011U 2W 18.001 RESF A 10KZ 0.006U 2W 19.001 RESF A 100KZ 0.06U 2W 20.001 RESF A 1000KZ 0.6U 2W 21.001 RESF A 10MZ 0.021U 2W 22.001 ASK+ R 22.002 DISP REMOVE THE TEST CABLE FROM THE 8600A 22.003 EVAL {Is the display flashing 18.888 with the 20Megohm LED lit?} 23.001 HEAD {DC CURRENT PERFORMANCE TEST} 23.002 DISP Select Range/Function 200uA DC 23.003 EVAL {DOES THE OPEN CIRCUIT READ -00.05 TO +00.05} 24.001 DISP CONNECT TEST CABLE TO MA AND LO TERMINALS 24.002 ASK- R 24.003 5100 200 190uA 0.21U 2W 25.001 ASK+ R 25.002 5100 2 1.9mA 0.0021U 2W 26.001 5100 20 19mA 0.021U 2W 27.001 5100 200 190mA 0.21U 2W 28.001 5100 2000 1900mA 2.1U 2W 29.001 DISP NOTE: Because the current measurements of ac and dc are 29.001 DISP made using the same shunt resistors; a check of ac cur- 29.001 DISP rent is not made. 29.002 HEAD {AC VOLTAGE PERFORMANCE TEST} 29.003 DISP DISCONNECT TEST CABLE FROM MA AND LO TERMINALS AND 29.003 DISP CONNECT TO V-OHM AND L0 TERMINALS THROUGH PRECISION 29.003 DISP 50-OHM FEED-THROUGH 29.004 WIDE 200 190mV 1.95U 70KH 2W 30.001 DISP REMOVE 50-OHM FEED-THROUGH AND RECONNECT TEST CABLE 30.002 5100 200 190mV 0.54U 500H 2W 31.001 5100 200 39mV 0.4U 30KH 2W 32.001 5100 2 1.9V 0.0041U 500H 2W 33.001 5100 2 0.9V 0.0021U 500H 2W 34.001 5100 2 1.9V 0.01U 50KH 2W 35.001 DISP REMOVE TEST CABLE FROM 8600A AND RECONNECT THROUGH 35.001 DISP PRECISION 50-OHM FEED-THROUGH 35.002 WIDE 2 1.9V 0.02U 100KH 2W 36.001 WIDE 2 1V 0.0055U 30H 2W 37.001 DISP REMOVE 50-OHM FEED-THROUGH AND RECONNECT TEST CABLE 37.002 5100 20 19V 0.041U 500H 2W 38.001 5100 20 19V 0.2U 50KH 2W 39.001 5100 200 190V 0.41U 500H 2W 40.001 5205 200 190V 0.55U 30KH 2W 41.001 5205 1200 1000V 3.1U 500H 2W 42.001 5205 1200 1000V 6U 20KH 2W 43.001 DISP THE 8600A VERIFICATION PROCEDURE IS COMPLETE END